VICI Metronics Dynacalibrators
Specification Model 120 Model 150 Model 230 Model 235 Model 340 Model 345 Model 450 Model 500 Model 505
CE Certification & NIST Traceable x x x x x x x x x
Temperature Accuracy ± 0.01°C ± 0.01°C ± 0.01°C ± 0.05°C ± 0.01°C ± 0.05°C ± 0.01°C ± 0.01°C ± 0.01°C
Temperature Range 5°C above ambient to 100°C Solid state proportional temp controller 30°C* to 110°C 5°C above ambient to 100°C 30°C* to 110°C 5°C above ambient to 110°C 30°C to 110°C 5°C above ambient to 110°C 30°C* to 110°C
Dimensions 10.5" W x 9.5" D x 13" H 15.4 cm wide x 38.1 cm deep x 17.7 cm high (6" W x 15" D x 7" H) "17"" wide x 23.75"" deep x 10.5"" high (43.2 cm x 60.3 cm x 26.7 cm)" 17" W x 23.75" D x 26.7" H "17"" wide x 23.75"" deep x 10.5"" high (43.2 cm x 60.3 cm x 26.7 cm)" 17" W X 23.75" D X 10.5" H "17"" wide x 23.75"" deep x 10.5"" high (43.2 cm x 60.3 cm x 26.7 cm)" 17” W x 23.75” deep x 10.5” high (43.2 cm x 60.3 cm x 26.7 cm) 17" W X 23.75" D X 10.5" H
Weight 25 lbs 10.5 lb. (4.76 kg) "W/ internal pump: 55 lbs (25.0 kg) Without internal pump: 3 lbs (19.5 kg)" 55 lb "W/ internal pump: 56 lbs (25.4 kg) W/o internal pump: 44 lbs (20.0 kg)" 22 lbs "W/ Internal pump: 59 lbs (26.8 kg) W/o: Internal Pump: 47 lbs (21.3 kg)" 74 lbs
Permeation chamber (s) Pyrex® permeation chamber (accommodates 18.5 cm total length Dynacal tube) Caustic chemicals (ex. HCl, HNO3, HF, & Cl) require Teflon® chamber 1
passivated stainless steel or optional Hastelloy C
1
passivated stainless steel or optional Hastelloy C
1
passivated stainless steel or optional Hastelloy C
1
passivated stainless steel or optional Hastelloy C
1
passivated stainless steel or optional Hastelloy C
1
passivated stainless steel or optional Hastelloy C

passivated stainless steel or optional Hastelloy C2
2
passivated stainless steel or optional Hastelloy C
Permeation chamber dimensions Glass 19.7 cm length x 2.1 cm diameter,
Teflon 19.7 cm length x 1.7 cm diameter
24 cm long by 1.6 cm (9.5” L by 0.875”) Accepts devices up to 23.5 cm long x 1.6 cm diameter (9.25” L x 0.62”) 24 cm x 1.6 cm 23.5 cm x 1.6 cm 24 cm x 1.6 cm 23.5 cm x 1.6 cm 23.5 cm x 1.6 cm
Flow
  • Model 120-1 0 to 1 lpm with certified settings at 200, 400, 600, 800 and 1,000 cc/min
  • Model 120-2 0 to 2 lpm with certified settings at 600, 1200, 1800 and 2400 cc/min
  • Model 120-4 0 to 4 lpm with certified settings 1, 2, 3 and 4 lpm
Flexible
Dilution flow capacity: 0-10 lpm with 15 certified settings
MFC**
1-20 SLPM *(depending on model)
1-20 LPM depending on model Two MFC**
1-20 SLPM *(depending on model)
1-20 LPM depending on model 1-20 LPM depending on model Two MFC**
1-20 LPM depending on model
Calibration modes Three Five (Auto or Manual) Span Out
User Interface
  • Permeation chamber access (screw cap)
  • Digital readout for chamber temperature
  • Flowmeter with flow control valve
  • Span and overflow outlet (1/4"" tube fitting)
  • Power switch with LED indicator light
  • Power receptacle for 110 VAC operation and recharging Solid state proportional temperature controller
  • Digital readout for set point and chamber temperature Teflon® mode select valve – Standby or Span
  • Heater switch with LED indicator light
Digital readout for set-point and chamber temperature New style controller with touch screen with mass flow control interface Touchscreen Touchscreen Digital thermometer (0-110°C) for precise chamber temperature readout
Long-life LED indicators
Touch screen interface for automatic or manual flow adjustments with chemical database
Feature Internal pump to supply scrubbed air to carrier stream Through-port Diverter valves
Carrier Flow Recommended range 100-1200 mL/min, must be supplied and measured externally Fixed at 80 cc/min or 175 cc/min depending on model Nominal flow rate 100 sccm Fixed at 80 cc/min or 175 cc/min depending on model Nominal flow rate 100 sccm Fixed at 80 cc/min or 175 cc/min depending on model Fixed at 80 cc/min or 175 cc/min depending on model Nominal flow rate 100 sccm
Automation *** Basic:
  • Provides continuous dilution
  • Maintains a constant carrier flow through the permeation chamber
Intermediate/ extended concentration range:
  • In the zero mode, scrubbed dilution flow is delivered to the outlet, allowing the end user to establish zero before sampling
  • Full range of mode capability
Dual chamber:
  • Two separate permeation chambers with independent temperature control systems
  • Chamber 1 and chamber 2 can run independently, or be used together to combine trace components
  • Solenoid valves allow the carrier flows to be switched from the dilution stream to a vent port, allowing chamber 1, char[1]

*or 2°C above ambient, whichever is higher
** MFC = Mass flow controller

*** For all automated models (235, 345, 505):

  • User sets either the flow rate or the concentration via touch screen
  • Required temperature and concentration or flow rate or set and controlled automatically
  • External gas source